    KLD Mono&Multiple Water Quality Analyzer
    On-line Analyzer for Ammonia and Nitrite and Nitrate Monitoring

    On-line Analyzer for Ammonia and Nitrite and Nitrate Monitoring

        KLD-C-NH3-N & NO3-N & NO2-N is an automatic on-line analyzer independently developed by KLD. Using imported CPU, highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.

    Hits:158
    Product Presentation

        KLD-C-NH3-N & NO3-N & NO2-N is an automatic on-line analyzer independently developed by KLD. Using imported CPU, highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.


    Features

    1.      Robust & Reliable

    Designed for on-line industrial and environmental applications. Complete separation between electronics and hydraulics allows long-term stable operations.

    2.      Precise Reagent Dosage

    CPU controlled photoelectric dosing system for accurate reagent dosage. Simple & steady structure ensures long-term service life.

    3.      Easy Installation

    Delivered after a long & successful series of factory tests, it is ready for installation and setup. To start monitoring just need to connect reagent tubes, sample tube, water tube, waste tube and power supply, then set up the parameters.

    4.      Automatic Calibration

    The analyzer performs calibration according to the calibration interval and type set up by users. If the result exceeds selected limits, the original calibration parameters will not be updated and an alarm signal will be outputted.

    5.      Automatic Monitoring

    Measuring interval can be freely set up by users. Between two measurements the analyzer remains in standby.


    Benefits

    ?  Automatic monitoring system gives alarm signal in abnormal condition. Easy and low maintenance; Low operating cost.

    ?  4-20mA analog signal output; standard RS232, RS485 digital signal output, MODBUS protocol.

    ?  New signal processor reduces signal fluctuation to ensure the accuracy and stability of data.

    ?  Highly precise peristaltic pump extends the service life of pump tubes.

    ?  Automatic restart when normal power supply after accidental power down.

    ?  Remote control with local or remote PC connection.

    ?  Low reagent consumption; short preparation time.

    ?  Automatic diagnosis function. Identification and alarm for reagent shortage.

    ?  Instrument operating signal output function.

    ?  3000 groups of data storage and USB download for analysis data and running log.

    ?  Easy operation, no special training is required.

    ?  Controllable running of water pump and free setup for the sampling time of the pump.

    ?  Security management function, automatic record and storage for all control operations. Read-only storage.

    ?  2-level operation & management authorization: The system administrator can have all system operation and general operators can only perform routine maintenance & operation.


    Measuring Principle

    NH3-N(OPA): The sample after proper filtration is pumped into the reactor, where OPA buffer solution are injected. When heated to some degree, the solution will react with ammonium ions then produce fluorescent reactants. The analyzer measures ABS value of the substances at the specific wavelength, then calculates ammonia concentration of the sample according to the stored calibration data.

        NO3-N: The sample after proper filtration is pumped into the reactor. Using copper sulfate as a catalyst, injecting hydrazine sulfate to convert nitrate into nitrite. Add SAA and NED to the sample. After fully mixing and reaction, the analyzer measures ABS value of red substances produced by reaction at the specific wavelength, then calculates sample concentration according to the stored calibration data.

        NO2-N: The sample after proper filtration is pumped into the reactor, where SAA and NED are injected. After fully mixing and reaction, the analyzer measures ABS value of red substances produced by reaction at the specific wavelength, then calculates sample concentration according to the stored calibration data.


    Specifications

    Measuring   Principle

    NH3-N

    NO3-N

    NO2-N

    Fluorimetric,   OPA Method

    Colorimetric,   NED+SAA Method

    Colorimetric,   NED+SAA Method

    Operating   Management

    Security   management and 2-level operation management function

    Measuring   Type

    Manual   or automatic with interval

    Measuring   Interval

    Programmable;   reverse control available by RS232

    Measuring   Time

    18   minutes

    16   minutes

    8 minutes

    Measuring   Range

    0.01~1/10/20/50mg/L,   other ranges available on request

    Calibration   Type

    Manual   or automatic with interval

    Precision

    ≤   5%

    Zero   Drift

    5%F.S

    Accuracy

    ±5%

    Range   Drift

    5%F.S

    Signal   Output

    Standard   4-20mA analog output,

    Standard   RS232/RS485 digital output, MODBUS Protocol

    Alarm

    High   limit alarm, low limit alarm, reagent shortage alarm, calibration error alarm   (standard)

    Sample   and Waste Delivery

    Pressure   free

    Reagent   Temperature

    10-30°C

    Reagent   Replacement

    3~4   weeks (depending on the ambient temperature)

    Ambient   temperature

    5-40°C

    Protection

    IP55

    MTBF

    720h

    Display

    Color   touch screen

    Power   supply

    220V   AC±10%; power 90W

    Weight

    60kg   (without reagents)

    Dimension

    1600×500×400mm(H×W×D)   (with the height of base)


    Optional Functions

    Dual-channel

    One analyzer to monitor on-line for 2 outlets.

    Reagent Cooling

    Longer reagent service life and less reagent replacing.

    Alarm Function

    DI water low limit alarm and waste liquid high limit alarm function.

    Automatic Dilution Function

    Automatic dilution function for high concentration sample.

    Mini Printer

    Print data after measurement.


      Tags:On-line Analyzer for Ammonia and Nitrite and Nitrate Monitoring
      Home | About | Job | Contact | Guess
      日本成a人片在线观看网址,www日韩中文字幕在线看,亚洲第一成人影院,亚洲欧美日韩一区高清中文字幕 绥江县| 鄯善县| 徐州市| 迁西县| 塔城市| 含山县| 呼和浩特市| 邵东县| 古丈县| 内江市| 游戏| 信宜市| 湘潭县| 鲁山县| 乌海市| 正镶白旗| 曲松县| 务川| 丘北县| 保靖县| 法库县| 鸡泽县| 四川省| 土默特左旗| 应用必备| 大方县| 滨州市| 敦煌市| 黔东| 洱源县| 济宁市| 彭水| 汝城县| 长宁区| 客服| 噶尔县| 高尔夫| 广汉市| 贵阳市| 宁德市| 米泉市| http://444 http://444 http://444 http://444 http://444 http://444